2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 198
1)
IF (and ONLY if) ONLY Valves #30 AND #26 ‘dots’ are blue AND all others are white:
(page 198).
2)
ALL other scenarios:
Proceed to
page 205
, procedure number
VE- 2.0.1 ONLY VALVES #30 AND #26 ‘DOTS’ ARE BLUE
a) At the distribution board (Figure below), unplug
Valve #26’s
connector.
b) Watch the Balancing Chamber Valve ‘dots’ for thirty (30) seconds. Do they begin to cycle between white
and blue?
Yes
The Balancing Chamber Valves cycle! Valve #26 OR its blue wire harness is bad.
No The Balancing Chamber Valve ‘dots’ REMAIN white! See procedure number
Figure 38 – Valves #26 AND #30
VE- 2.0.2 VALVE #26 UNPLUGGED AND THE BALANCING CHAMBER VALVES REMAIN WHITE
a) At the distribution board, Figure above, unplug
Valve #30’s
connector.
b) Watch the Balancing Chamber Valve ‘dots’ for up to thirty (30) seconds. Do they begin to cycle between
white and blue?
Yes
The Balancing Chamber Valves begin to cycle! Valve #30 OR its blue wire harness is bad.
No a) CAREFULLLY return all valve connectors to the distribution board;
b) THREE (3) possible bad components, see the Component List below:
Component List
Turning the machine off first, swap in each component, one at a time, returning to Dialysis
Program in between, until
VERR
remains = 0!
1)
Actuator-Test Board;
2)
Actuator-Test board Cable;
3)
Distribution board
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...