2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 413
FIL- 1.0.3 DEAERATION PRESSURE OKAY (2)
Because the machine was off, [
Conductivity
] will be unstable! The following procedure requires three (3)
minutes during when [
Conductivity
] MUST stabilize to between 13.0 and 14.4 mS!
a)
OPEN THE SHUNT DOOR AND LEAVE IT OPEN TILL INSTRUCTED!
b) Call debug screen 0. WITHOUT LOOKING AWAY watch
Flow
Error
for
three (3) minutes
! If EVER = 1, even just once,
indicates a masked Flow Error! TWO (2) possible scenarios:
1)
IF (and ONLY if) Flow Error ALWAYS = 0!
(page 413).
2) IF Flow Error EVER = 1:
Proceed to
page 20
SECTION 1 - FLOW ERRORS IN DIALYSIS
FIL- 1.0.4 ISOLATE ‘OUT OF BYPASS’ CIRCUIT
a) From debug screen 0, Figure right, TWO (2) IMPORTANT
checks:
CHECK #1: Pre Dialyzer Temperature
MUST BE stable
*
BETWEEN 35.0 and 39.0° C!
CHECK #2:
Pre Dialyzer
Conductivity
MUST BE stable
*
BETWEEN 13.0 and 14.4 mS!
*
Stable = NOT changing more than 0.1 per minute!
If
Temperature
is about 37.0° C and
Conductivity
about 14.0 mS but either one bounces’ more than 0.3 may indicate a large air leak
that must be located and repaired!
(page 413).
FIL- 1.0.5 STABLE CONDUCTIVITY / ISOLATE OLC
A bad On Line Clearance (OLC) component may cause Filling Programs EVEN if OLC is not activated or
used! From debug screen 0:
a)
Figure right
, if the
POST Dialyzer Cond
symbol
appears see part b. If NOT skip to procedure number
(page 414).
b) Are
PRE AND POST
Dialyzer Conductivity within 0.4
mS of each other (Yes or No)?
Yes
PRE AND POST Conductivity within 0.4
mS of each other! Proceed to
page 416
procedure number
No PRE AND POST Conductivity ARE NOT within 0.4 mS of each other! See procedure number
(page 414).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...