2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 578
Yes
“HEATER RELAY TEST FAIL” reoccurs! See procedure number
(page 578).
No
“HEATER RELAY TEST FAIL” did NOT reoccur! See procedure number
page 583
HEATER RELAY- 1.0.3 ISOLATE HEATER OFF LOGIC
a) Turn the machine OFF!
b)
IMPORTANT!
Turn the HEATER Switch off!
c) Turn the machine on AND r
eturn to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘CONFIRM’).
d) Does
“
HEATER RELAY TEST FAIL” reoccur?
Yes
“HEATER RELAY TEST FAIL” reoccurs! ENSURING the Heater Switch was off, see
procedure number
No
“HEATER RELAY TEST FAIL” did NOT reoccur! See parts a AND b below:
a)
IMPORTANT!
Turn the Heater Switch on!
b) CAREFULLY repeat (ABOVE) procedure number
(page 577)
HOWEVER, if you return here see procedure number
).
HEATER RELAY- 1.0.4 HEATER SWITCH OFF BUT “HEATER RELAY TEST FAIL” REOCCURS
a)
IMPORTANT!
Leave the Heater Switch off until the problem is located!
b) Turn the machine off and swap in a known good Power Logic Board. To LOCATE the board refer to
(page 9).
c) R
eturn to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘CONFIRM’). If (and ONLY if) “HEATER
RELAY TEST FAIL” reoccurs see procedure number
(page 578). If it does not
reoccur the new Power Logic board solved the problem!
HEATER RELAY- 1.0.5 HEATER RELAY TEST FAIL REOCCURS
NOTE:
The previous Power Logic Board is probably okay.
a) Turn the machine off. THREE (3) possibilities (see COMPONENT LIST below). Swap each in, one at a
time then in between, see part b to see if the new component fixed the problem.
COMPONENT LIST: 1)
Power Control board (inside the power supply);
2)
Functional Board (to prevent
“Cond Offset Failure”, place the machine into
T and C Mode
(refer to
OPERATING MODES
, page
Error! Bookmark not defined.
));
3)
Motherboard.
b) Turn the machine on and r
eturn to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘CONFIRM’). If
“
HEATER RELAY TEST FAIL” reoccurs swap in the next component in the list
“
HEATER RELAY TEST
FAIL” does NOT occur.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...