2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 199
VE- 3.0.0 ALL BALANCING CHAMBER VALVES CYCLING BETWEEN WHITE AND BLUE
a) Call debug screen 1.
b) NOTE
VERR’s
value for future reference. If
VERR
EVER reaches 255 return to procedure number
(
page 196
) to reset it.
c) Figure right, if (and ONLY if) a “Dial Valve Failure” OR
“Act Byp Valve Fail” banner EVER appears proceed to
page 652
d) Watch
VERR
until if it increases above what was noted OR for ten (10) minutes whichever comes first!
TWO (2) possible scenarios:
1)
IF (and ONLY if) VERR increases above what was noted in part b:
Return to ABOVE
(
page 196
).
2)
IF VERR does DOES NOT increase:
(page 199).
VE- 3.5.0 ISOLATE VALVE #43
a) Remove the resistor plug from the Air Sensors distribution board position but
DO NOT
return the Air
Sensor’s connector yet!
b) Call debug screen 0. Chamber #69 MUST say “Air”.
c) Call debug screen 1. Watch
VERR
until if it increases above what was noted in
procedure number VE- 3.0.0, part b OR for five (5) minutes whichever comes
first! Does
VERR
increase?
Yes
VERR
increases! See ABOVE procedure number
(page 197).
No
VERR
does DOES NOT increase!
Return the Air Sensor’s connector to it distribution board
position. A valve error is not occurring at this time! If a Flow Error reoccurs proceed to
page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...