2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 232
T- 1.0.20 FLOW THROUGH THE EXTERNAL INDICATOR
a) Leaving Dialysate Flow on, install a Temperature meter (°C)
into the dialysate lines.
IMPORTANT! Figure right, Flow
MUST be bottom to top!
b)
CLOSE THE SHUNT DOOR!
c) ENSURING the flow indicator’s ‘bob’ is moving, see procedure
(page 232).
T- 1.0.21 ISOLATE ‘ACTUAL’ (MEASURED) TEMPERATURE
Is the meter’s ‘measured’ temperature LESS THAN 35.1° C?
Yes
Temperature is LESS THAN 35.1° C! See procedure number
(page 234).
No Temperature is MORE THAN 35.1° C! See parts a THROUGH e below:
a) Call debug screen 1.
b)
Do NOT reset alarms!
c) Press and hold the “1” key for five (5) seconds.
d) Allow FORTY-FIVE (45) seconds BEFORE continuing to part e.
e) Watch the meter for one (1) minute. TWO (2) possible scenarios:
1) IF between 35.1° C and 38.9° C AND STABLE i.e. does NOT change more than
0.2° C per minute:
Proceed to
page 235
, procedure number
2) IF UNSTABLE i.e. changes more than 0.2 °C per minute:
Allow ninety (90) more
seconds THEN, ENSURING the ‘bob’ is moving up and down
*
, watch the meter for one (1)
more minute. Does it STOP changing more than +/- 0.2 °C (Yes or No)?
*
Note:
If the ‘bob’ has stopped moving call debug screen 0. If
Flow Error
= 0 return to
(page 229).
Yes
Measured temperature STABLIZES. If (and ONLY if) between 35.1 and 38.9º C proceed
to
page 235
, procedure number
. If not between 35.1 and 38.9°C return to
(page 232).
No
Measured temperature continues to be unstable. Watch the meter for five (5) more
minutes. THREE (3) possible scenarios 1) or 2) or 3) next page:
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...