2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 372
CO- 8.7.0 ISOLATE BALANCING CHAMBER DIAPHRAGMS
a) BEFORE continuing to part b,
NOTE
this page and procedure number in case you need to return here
then proceed to
page 535
SECTION 19 - TESTING FOR A LEAKING BALANCING CHAMBER
b) If a leaking balancing chamber diaphragm was NOT located in part a return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
(page 372).
CO- 8.8.0 VERIFY VOLUMETRIC SYSTEM
A 100 ml
graduated cylinder (preferably glass) is REQUIRED!
CAUTION!
Using a larger cylinder
results in gross error!!
If a CALIBRATED scale is available use it (1 gram = 1 ml)!
a)
WITHOUT turning flow off
, remove the tubing from the ‘To Drain’
(
BOTTOM*
) nozzle at the REAR OF THE MACHINE.
*
Do NOT measure from the drain tubing to prevent gross error!!
b)
Allow ten (10) drain ‘pulses’ to
ENSURE stability!
c)
ACCURATELY,
collect two
consecutive drain ‘pulses’ into
the 100 ml cylinder.
d) Per the Figure right, read the
bottom of the meniscus curve.
e) Call debug screen 3.
f)
BCV
(upper right,
B
alancing
C
hamber
V
olume) was entered during the last calibration
and is the volume of both chambers (
example:
603 = 60.3 ml).
g) Does
BCV
AND the measured volume times (x) 10 match?
Yes
BCV
= measured volume! See procedure number
(page 373).
No
a)
Read before performing!
Perform BALANCING CHAMBER VOLUME CALIBRATION,
per the CALIBRATION PROCEDURES booklet but enter your above measured volume.
b) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM)’!
c) Allow ten (10) minutes BEFORE continuing to part d.
d) Is the external meter’s conductivity /- 0.4 mS of TCD?
Yes
Conductivity /- 0.4 mS of TCD! The balancing chamber was out of
calibration. Do NOT continue!
No Conductivity IS NOT /- 0.4 mS of TCD! See procedure number
(page 373).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...