2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 311
T- 7.6.0 METER TEMPERATURE IS STABLE AT 37°C / CHECK NTC #3 STABILITY
The screen’s
[Monitor Reference]
data box should be gray AND at this point it does not matter if it = 37° C!
WITHOUT LOOKING AWAY, watch it one (1) minute noting its highest and lowest value. TWO (2) possible
scenarios 1) or 2) below:
1)
IF (and ONLY if) [Monitor Reference] changes more than +/- 0.2° C:
ENSURING the external
meter is STABLE between 36.9 and 37.1° C, proceed to
page 314
, procedure number
2) IF [Monitor Reference] does NOT change more than +/- 0.2° C:
See parts a AND b below:
a) ‘Sharply’ press ‘CONFIRM’ to turn the
[Monitor Reference]
data box pale yellow / white.
b) Activate
[Monitor Reference]
,
it turns bright yellow. Without pressing CONFIRM until instructed can
you adjust
[Monitor Reference]
to match the external meter?
Yes
[Monitor Reference]
should be bright yellow
AND between
36.9 and 37.1° C. See
No Swap in the listed components (see COMPONENT LIST below), one at a time, with known
good, in between returning to (ABOVE) procedure number
(page 305) to test each
new component.
COMPONENT LIST: 1)
NTC #3 (to LOCATE NTC #3 refer to
2)
Functional Board.
T- 7.6.2 TEMPERATURE CHECKS / SLOW FLOW CALIBRATION?
CAREFUL HERE! This procedure determines if a ‘Slow Flow’ Temperature calibration is necessary:
a) ‘Sharply’ press ‘CONFIRM’!
[Monitor Reference]
MUST return to pale yellow / white!
b) ‘Sharply’ press ‘CONFIRM’ again. TWO (2) possible scenarios:
1)
IF the screen says “9. Connect an external temperature meter…”:
A ‘Slow Flow’ calibration is
required! See procedure number
(page 312).
2) IF the screen says “Calibration saved. Press CONFIRM”:
If the procedures were done correctly
the
[Monitor Reference]
data box is GRAY. Does it continue to read between 36.9 and 37.1° C?
Yes
The GRAY data box = between 36.9 and 37.1° C. This completes the calibration
HOWEVER, if you started out troubleshooting a “TEMP OVER 95 DEGREES” alarm in
Heat Disinfect return to procedure number
(page 247). If not, return to Dialysis
Program.
No The GRAY data box does NOT = between 36.9 and 37.1. Swap in the listed components
(see COMPONENT LIST below), one at a time, with known good then return to (ABOVE)
procedure number
(page 305) to test each new component.
COMPONENT LIST:
1)
NTC #3 (to LOCATE NTC #3 refer
2)
Functional Board.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...