2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 382
CO- 10.0.26 TROUBLESHOOTING POST DIALYZER TEMP / POST ISOLATE NTC #44
a) Return NTC #44’s connector to distribution board position “X44, NTC-POST”.
b) From the Home screen, set [Dialysate Flow] to 500 ml/min and press ‘CONFIRM’.
c) Allow the Home screen’s [
Temperature]
display to stabilize between 35.5 and 38.5° C.
d) Call debug screen 0. Is
Post Dialyzer Temp
more than 33° C?
Yes
Post Temp more than 33! See procedure number
(page 382).
No Post Temp is less than 33! Post Dialyzer Temp Sensor NTC #44 may be bad. To
LOCATE NTC #44 refer to
(page 378).
CO- 10.0.28 TROUBLESHOOTING POST DIALYZER TEMP SENSOR / CALIBRATE TEMP
a)
Read parts b through f before performing!
b) Perform the TEMPERATURE CONTROL calibration per the CALIBRATION PROCEDURES booklet.
c) After the calibration, return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
d) From the Home screen, allow [
Temperatur
e] to stabilize between 35.5 and 38.5° C.
e) Open the shunt door and allow one (1) minute.
f) From debug screen 0, is
Post Dialyzer Temp
now /- 0.6° C of
Pre Dialyzer Temp
?
Yes
Post Temp within 0.6 of pre Temp. Problem solved! The Post Dialyzer Temp Sensor
NTC #44 was out of calibration.
No Post Temp is NOT within 0.6 of pre Temp. Post Dialyzer Temp Sensor NTC #44 may be
bad. To LOCATE NTC #44 refer to
(page 378).
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...