2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 74
F- 7.0.9 ISOLATE FLOW FROM THE BALANCING CHAMBERS
a) Figure below, clamp and remove the bottom tubing at Pre-Cell #7
b) Direct the tubing into the cylinder.
c)
IMPORTANT!
Remove the clamps from Pre-Cell #7 AND valve #26!
d) Measure flow, from the tubing for one (1) minute. 800 ml or more collected?
Yes
800 or more!
A)
Clamp and reattach the tubing;
B)
No
Less than 800 ml! See parts a AND b below:
a)
Return valves #30 and #37 to their respective distribution board positions.
b)
Read before performing!
CAREFULLY repeat from (ABOVE) procedure number
(
page 65
) but if (and ONLY if) you return here THREE (3) possible bad components:
1)
Bad Actuator-Test Board
1
OR;
2)
Bad Actuator-Test board cable OR;
3)
Multiple
restricted balancing chamber valves
2
#31, #33, #35 and #37.
1
To LOCATE the Actuator-Test board refer to
(page 9)
2
To LOCATE valves #31, #33, #35 and #37 refer to
(page 68)
F- 7.1.0 ISOLATE DIASAFE
®
FILTER RESTRICTION
a)
IMPORTANT!
Remove the clamp from the Pre-Cell #7’s tubing.
b) Using a flashlight, check for restrictions (possibly bio-growth), through the tubing, to and from the
balancing chamber valves and the DiaSafe
®
filter. If no restrictions are located, replace the DiaSafe
®
filter per procedure.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...