2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 528
Table 14 – Leak Test Scenarios
Observed Scenario:
Your Response:
If ALL valves PASS
Return to the NOTED procedure that brought you here
If ONLY one valve fails
(page 532)
If ONLY Valves #31 and #35 fail
(page 532)
If ONLY Valves #32 and #36 fail
(page 532)
If ONLY Valves #33 and #37 fail
(page 532)
If ONLY Valves #34 and #38 fail
(page 532)
ALL OTHER scenarios
Turn the machine OFF then see procedure number
(page 528)
AUTOTEST- 1.0.0 IOS FAILED / POSITIVE PRESSURE LEAK TESTS
ANY external leak, however small, MAY be causing IOS failures and MUST be repaired!
a) Place the machine into RINSE PROGRAM.
b) Using a flashlight, THROUGHLY check for external leaks.
c) Check the
ENTIRE
length of the external dialyzer lines for leaks (i.e. dialyzer quick connectors, external
flow indicator, connections at the rear of the machine and (if equipped) the sample port.
d) Check the DiaSafe
®
filter tubing AND its housing for external leaks!
e) If an immediate leak is NOT located allow two (2) minutes then repeat parts a THROUGH d! If still NO
(page 528).
AUTOTEST- 2.0.0 NO EXTERNAL LEAKS LOCATED / CHECK UF PUMP
The UF Pump creates negative pressure for the Valve Leak Test. If it is malfunctioning multiple valves fail:
a) Enter Service Mode
→ Calibrate Hydraulics → UF Pump Volume.
b) Place the dialyzer connectors into a bucket of water on the floor.
c) Close the shunt door.
d) Figure right, remove the Ultrafiltrate Output Sample Connector from
its port and hold it up as seen in the Figure!
e) Press ‘CONFIRM’. The screen changes.
f) Press the ‘Prime’ key. IGNORE five (5) strokes then watch five (5). Does the UF Pump deliver ‘strong
pulses’, through the Fluid Sample Connector, that squirt out at least six (6) feet?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...