2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 363
CO- 7.0.0 TROUBLESHOOTING COND CELL #7 CIRCUIT
a)
IMPORTANT!
Turn the machine OFF!
b) Figure right, place the
274 Ω
resistor plug, from the
TWO-
RESISTOR SET
, into
Conductivity Cell’s #7
distribution board position, “X7,
COND”.
c) Install a conductivity meter into the dialysate lines to measure ‘actual’ conductivity subsequently!
d) Place the machine into Service Mode
→ Calibrate Sensors → Cond Cells. Follow the screen’s
instructions through:
“4. Waiting until the Conductivity value is stable.”
e) ALLOW six (6) minutes before continuing!
f) Verify measured Conductivity is between 13.0 and 14.3!
g) If Cond Cell #7’s reading stabilizes,
“5. Conductivity value is stable….”
appears?
Yes
“5. Conductivity value is stable…”
Do NOT continue with the calibration! Cond Cell #7*
may be bad. *To locate Cell #7 refer to
(page 344).
(page 364).
LEFT BLANK INTENTIONALLY