2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 342
CO- 2.0.9 ‘BOB MOVING’ / MEASURE TEMPERATURE
a) WITHOUT turning [Dialysate Flow] off, install a
Temperature / Conductivity meter into the dialysate lines.
IMPORTANT!
Figure right, flow through the meter MUST
be from bottom to top!
b)
IMPORTANT! Close the shunt door!
c) Set the meter to measure temperature (° C).
d) Call debug screen 1.
e) Press and hold the “1” key for five (5) seconds.
f) Allow
two (2) full minutes
BEFORE
continuing to part g!
g) ENSURING the flow indicator’s ‘bob’ is moving up and down, is the external meter stable between
35.5 and 38.5° C?
Yes
Between 35.5 and 38.5° C! See procedure number
(page 342).
No
NOT between 35.5 and 38.5!
Proceed to
page 302
, procedure number
CO- 2.0.10 TEMPERATURE BETWEEN 35.5 AND 38.5 / MEASURE CONDUCTIVITY
a) Set the meter to measure Conductivity (mS).
b) Measured Conductivity is compared to
TCD
as seen in debug screen 1’s,
C TCD
(lower left column). Example: 1360 = 13.60 mS.
c) ENSURING the flow indicator’s ‘bob’ is still moving up and down, based on the external meter’s
(measured) reading versus
TCD
, THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) measured conductivity is within 0.4 of TCD AND stable i.e. does NOT
change more than +/- 0.15 per minute:
See procedure
(page 343).
2)
IF (and ONLY if) measured conductivity
IS NOT
within 0.4 of TCD:
Proceed to
page 366
3)
IF measured conductivity IS MOSTLY within 0.4 of TCD BUT unstable i.e. changes more
than 0.15 per minute:
See parts a AND b below:
a) If removed from a previous procedure, return
Conductivity Cell #7’s connector to distribution
board position “X7, COND” (Figure right).
(page 350).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...