2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 166
CLEAN- 2.0.0 LESS THAN 300 ML PER MINUTE COLLECTED / ISOLATE RESTRICTOR #48
This procedure attempts to clear a DEAERATION PUMP air lock. First Restrictor #48 is checked to see if it
is plugged:
a) Turn the machine off then back on.
b) Allow the “Select Program” banner to appear but
DO
NOT
select any screen buttons! “Select Program”
remains up until
INSTRUCTED OTHERWISE!
c) Figure right, a
60 ml syringe
, with its plunger pushed in
fully, is REQUIRED!
Using a smaller syringe may cause
error!
d) Attach the syringe to the Deaeration Pump’s CLEAR (INPUT) tubing removed earlier.
e) PULL on the plunger to the end of the barrel! If Restrictor #48 is plugged you will feel very strong
resistance and may not be be able to pull the plunger all the way out! Very strong resistance felt?
Yes
Very strong resistance! Restrictor #48 is plugged with debris
1
. Replace the tubing segment
that contains Restrictor #48*. *To LOCATE Restrictor #48 refer to the Figure, next page!
1
Debris indicates FIVE (5) possibilities:
1)
Excessive O-ring lubrication;
2)
Inadequate acid
cleaning (bicarbonate precipitate);
3)
Inadequately filtered incoming water;
4)
Degraded Heater;
5)
Degrading Deaeration Pump Head.
NOTE:
After the repair (i.e. good loading pressure achieved) see procedure number
(page 171).
No Very little resistance! See part A below:
A)
Remove the syringe and, from the tubing, MEASURE flow for one (1) minute. More than
130 ml collected?
Yes
More than 130 ml collected! See procedure number
No Less than 130 ml collected! Perform parts B through D below.
B)
Push the plunger back into the barrel then
reattach the syringe to the Deaeration Pump’s
tubing.
C) Pull on the plunger to the end of the barrel!
D) Repeat parts A through D up to
TEN (10) TIMES
OR until you MEASURE more than 130
ml in one minute?
Yes More than 130 ml per minute! See procedure number
167).
No After ten (10) attempts you CANNOT achieve more than 130 ml per minute!
Proceed to
page 169
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...