2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 386
CO- 10.0.41 PREPARE TO ANALYZE PRE AND POST CONDUCTIVITY STABILITY
a) Install a ‘dummy chamber’ in the Level Detector module and RESET ALL alarms!
b) Watch the screen for thirty (30) seconds. If a “Filling Program” banner occurs, allow thirty (30) seconds
before continuing!
NOTE:
Multiple “Filling Programs” occurrences indicate an air leak that
MUST
be located and
repaired or else OLC tests cancel!!
c) If screen says “Standby For Test” OR the screen says “Test:…” allow the tests to finish THEN allow
three (3) minutes BEFORE continuing to procedure number
CO- 10.0.42 ISOLATE FLOW PUMP DIGITAL CONTROL (FLWP)
a) Call debug screen 1.
b) Press the “1” key for five (5) seconds.
c) Allow forty five (45) seconds BEFORE continuing to part d.
d) Call debug screen 0. Without resetting alarms, ENSURING
Flow Error
REMAINS = 0 ALSO watch
FLWP
for two (2) minutes. TWO (2)
possible scenarios:
1)
IF (and ONLY if) FLWP continually changes more than +/- 2 (two):
Proceed to
page 44
, procedure number
2) IF FLWP does NOT change more than +/- 2 (two):
Reset ALL alarms
(page 386).
CO- 10.0.43 PRE CONDUCTIVITY STABILITY
a) Call debug screen 5.
Pre Dev val
(right column) is the conductivity difference between
now and one (1) minute ago from Pre-Conductivity Cell #7. If no alarms it changes
continually!
b) After no more than two (2) minutes, if not already,
Pre Dev val
should
fall to less than 135. TWO (2)
possible scenarios:
1)
IF (and ONLY if) Pre Dev val is OR falls to less than 135:
See procedure number
(page 387).
2) IF Pre Dev val NEVER falls below 135:
Pre Cond is unstable!
ENSURING no alarms occurred
see parts a AND b below:
a) Per the Figure (next page), using a flashlight and WITHOUT LOOKING AWAY, for one (1)
minute, if air bubbles are flowing through
Pre Dialyzer Cond Cell #7’s
tubing locate and repair
the air leak!
b) If (and ONLY if) absolutely sure of no air, proceed to
page 391
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...