2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 43
F- 3.3.0 GOOD FLOW PUMP PRESSURE
a) Return the connectors to acid and bicarbonate.
b) Return to Dialysis Program (“Select Program”
→ ‘Dialysis’ → ‘CONFIRM’)!
c) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’!
d) Allow [
Conductivity]
to increase to more than 13.0 mS!
e) TWO (2) possible TMP window scenarios below:
1)
IF (and ONLY if) the TMP window is NOT red:
2) IF the TMP window is red (TMP alarm):
See parts a THROUGH c below:
a) Press and release the ‘Reset’ key then immediately press and hold it for three (3) seconds.
b) Allow thirty (30) seconds.
c) If a TMP alarm reoccurs, repeat parts a and b up to twice more BEFORE continuing to
F- 3.6.0 ANALYZE FLOW ERROR
a)
Allow thirty (30) seconds BEFORE continuing to part b!
b) Call debug screen 6. WITHOUT LOOKING AWAY, watch
BC Switch
(middle
column)
for five (5) full minutes OR until if it EVER = 897 or more, even just once.
THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) REMAINS ALWAYS = 897 or more:
Return to (ABOVE) procedure number
page 30).
2)
IF (and ONLY if) NEVER = 897 or more:
The Flow Error must be VERY intermittent!
(page 44) may locate it!
3)
IF EVER cycles to 897 or more but does NOT remain 897:
See parts a THROUGH e below:
a) Replace the DiaSafe
®
filter preferably with one from another machine that is not exhibiting
Flow Errors.
b) Place the machine into RINSE for five (5) minutes to prime the filter.
c) Connect to concentrate and return to Dialysis Program
(‘Dialysis’ → ‘CONFIRM’).
d) From the Home screen, allow [
Conductivity
] to increase to more than 13.0 mS.
e) Call debug screen 6 to watch
BC Switch
for five (5) full minutes OR until if it EVER = 897 or
more! THREE (3) possible scenarios next page:
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...