2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 569
LEAKING- 7.0.0 “SAMPLE BAG CONNECTED?” / ISOLATE DIALYSATE SAMPLING OPTION
A leak was occurring in a Cleaning Program (Rinse, Heat Disinfect, etc.) and you saw the “Sample Bag
Connected?” banner on the screen.
a) Turn the machine on. When “Press CONFIRM For Service Mode” appears press ‘CONFIRM’. The screen
says “Machine In Service Mode”.
b) Select the screen’s ‘Options’ button.
c) Select the screen’s ‘Hardware Options’ button.
d) Does the “Yes” box at
Dialysate Sampling
(Figure right) have a blue ‘X’
in it?
Yes
Dialysate Sampling
(page 569).
No
Dialysate Sampling
= “No”! This is NORMAL and is not causing the leak! Turn the machine
OFF then return to (ABOVE) procedure number
LEAKING- 7.1.0 DIALYSATE SAMPLING = “YES”
a) Select
Dialysate Sampling
.
b) Place the ‘X’ in the “No” box and press ‘CONFIRM’.
c) ENSURE the “No” box at
Dialysate Sampling
now has a blue ‘X’ in it!
d) Reattach and all ‘blown off’ tubing but
LEAVE
the Fluid Sample Connector in the bucket!
e) Turn the machine OFF then back on!
f) From the ‘Select Program’ screen select RINSE then press ‘CONFIRM’. The “Sample Bag Connected”
banner should NOT appear!
g) With the machine in RINSE Program
NOTE:
Squirting from the Fluid Sample Connector is NORMAL =
UF Check Valve #63 is oriented correctly!
h) If (and ONLY if) an active leak reoccurs turn the machine OFF then return to (ABOVE) procedure
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...