2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 300
T- 6.6.0 VERIFY INCOMING WATER SOURCE
Is the machine connected to the:
1) SAME water source as other currently RUNNING machines
that
have good stable temperature OR
2) Its OWN water source
, possibly a portable RO. TWO (2) possible
scenarios:
1)
IF (and ONLY if) connected to the SAME water source as other machines:
See procedure
2) IF connected to its OWN water source:
Using a calibrated temperature meter check the incoming
water temperature (i.e. from the RO). If NOT between 10 and 33° C (50 and 86° F) this may be the
problem! If (and ONLY if) between 10 and 33° C see procedure number
(page 300).
T- 6.8.0 CHECK / ROTATE HEATER
a)
Turn the machine OFF!
b) Figure right, loosen the heater mounting
bracket and pull the heater out.
AVOID
touching it!
c) Replace the heater if badly discolored or if
corrosion or pitting is located.
d) Rotate the heater 180° and reinstall it.
Tighten BOTH* bracket screws evenly!
The heater
MUST
be positioned perfectly
vertical to maintain proper flow geometry.
*
A missing screw or mounting nut may be the problem!
e) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
f) Allow ten (10) minutes before continuing to part g.
g) If
Temperature
instability continues, changes more than 0.2 °C per minute, see procedure number
(page 300).
T- 6.9.0 UNSTABLE TEMPERATURE CHECKS
Other procedures, in different Sections of the Guide, are performed next.
IMPORTANT!
NOTE
this page
and procedure number as you may prompted to return to here. See part a below:
a) BEFORE continuing to part b,
page 535
SECTION 19 - TESTING FOR A LEAKING
b) If leaking diaphragm was not located in part a, turn the machine OFF!
Parts c and d next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...