2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 200
VE- 4.0.0 FOUR BALANCING CHAMBER VALVES REMAINING BLUE, FOUR REMAINING WHITE
NOTE:
One of the valve circuits whose ‘dots’ are blue is the problem. These procedure isolate between
them:
a) Turn Dialysate Flow off (lamp above the Flow on/off button blinks).
b) Return to screen 0. If flow is off all eight (8) Balancing Chamber Valves remain white.
c) Call debug screen 1!
NOTE
VERR’s
value for reference later. If
VERR
EVER reaches 255 return to
(
page 196
) to reset it.
d) Watch
VERR
until if it increases above what you noted in part d OR for five (5) minutes whichever
comes first! TWO (2) possible scenarios:
1)
IF (and ONLY if) VERR increases above what was noted in part d:
Leaving Dialysate Flow
OFF,
proceed to
page 205
, procedure number
2)
IF VERR does NOT increases above what was noted in part d:
This indicates a problem with a
Balancing Chamber Valve circuit. See procedure number
(page 200).
VE- 4.0.1 ISOLATE VALVES
a)
Turn Dialysate flow on (flow on/off lamp stops blinking).
b) Call debug screen 0.
c) Allow the Balancing Chamber Valve ‘dots’ to STOP cycling i.e. four (4) REMAINING blue! Four (4)
REMAINING white! This may take a while so be patient!
d) Per the Figures below, TWO (2) possible scenarios:
1)
IF (and ONLY if) V32, V33, V35 AND V38 are REMAINING blue:
The problem is in Balancing
Chamber Valve Cycle 1! See procedure number
(page 201).
2)
IF V31, V34, V36 AND V37 are REMAINING blue:
The problem is in Balancing Chamber Valve
Cycle 2! Proceed to
page 203
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...