2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 90
F- 8.2.0 ANALYZE ACFS
a) Call debug screen 0. Figure right, look at
ACFS
(Figure right) THEN locate
its value in
COLUMN
1 of the Table below.
b) Call debug screen 10.
c) Look at
ACFS
(right column, bottom) THEN locate its
value in
COLUMN 2
of the Table below next to screen 0’s
ACFS
value seen in part a.
d) Respond according to the Table’s
COLUMN 3,
based on BOTH
ACFS
readings:
COLUMN 1
Screen 0’s
ACFS
COLUMN 2
Screen 10’s
ACFS
COLUMN 3
Your Response:
Between 3.0 and 6.0
(Good)
Between 4.0 and 5.8
(Good)
BOTH
ACFS
values are good! The CFS
circuit is good! See procedure number
(page 91)
Between 3.0 and 6.0
(Good)
Less than 4.0 OR more than 5.8
(Bad)
Screen 0’s
ACFS
is good; Screen 10’s
ACFS
is bad
THREE (3) possible bad components
1
:
1)
Sensor Board OR;
2)
Functional Board
OR;
3)
Mother board connection between
the Functional and Sensor Board
Less than 3.0 OR more
than 6.0
(Bad)
Between 4.0 and 5.8
(Good)
Screen 0’s
ACFS
is bad; Screen 10’s
ACFS
is good
THREE (3) possible bad components
2
:
1)
Actuator-Test Board OR;
2)
Bad Sensor
Board OR;
3)
Motherboard connection
between the Actuator-Test and Sensor
Board
Less than 3.0 OR more
than 6.0
(Bad)
Less than 4.0 OR more than 5.8
(Bad)
BOTH bad! Proceed to
page 103
1
Referring to OPERATING MODES (page
Error! Bookmark not defined.
), to prevent “Cond Offset
Failure”, place the machine into
T and C Mode
THEN , with the machine off, swap these components in, one at a time, and in between,
when screen 10’s
ACFS
goes to between 3.0 and 6.0 the last component swapped in is the culprit.
2
Referring to OPERATING MODES (page
Error! Bookmark not defined.
), to prevent “Cond Offset
Failure”, place the machine into
T and C Mode
THEN, with the machine off, swap these components in, one at a time and in between,
when screen 0’s
ACFS
goes to between 3.0 and 6.0 the last component swapped in is the culprit.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...