2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 398
Figure 63 – Distribution Board / Hydraulics Top View (Pumps)
EOS- 2.0.0 ACID SYSTEM PROBLEM / ISOLATE THE ACID PUMP
This procedure uses the known good bicarbonate pump to check the acid pump.
A)
IMPORTANT! Turn the machine off to prevent damage!
above, unplug the acid pump cable from distribution board position “P16, CONC-P”.
CAUTION!
Avoid the Heparin Pump connector at position “P18”!
C) Plug the bicarbonate pump’s cable into the acid pump’s position “P16, CONC-P”.
CAUTION! The
connector should insert easily if the pins are okay!!
D) If the problem was occurring in a Cleaning Program return to it. If not return to Dialysis Program.
E) Call debug screen 0. If debug does not appear press ‘Escape’ then call debug 0.
F) Allow thirty (30) seconds THEN,
ignoring the BIC symbol
, watch the CONC symbol for up to two (2)
minutes:
1) EVER pink
(solid or blinking)?
OR
2) Remaining ALWAYS white
?
1)
IF ever pink:
The acid pump is probably okay. See procedure number
(page 400).
2) IF remaining white:
See parts a THROUGH c below:
a)
IMPORTANT! Turn the machine OFF to prevent damage!
b) Return the bicarbonate pump’s cable to position “P17, BIC-P”.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...