2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 156
c) ENSURE the ‘to drain’ tubing is not kinked AND (if
used) the ‘Quick connector’ is attached PROPERLY
to the station.
d) Figure right,
till INSTRUCTED OTHERWISE
,
direct the Fluid Sample Connector into a bucket on
the floor!
e) Below the machine will be returned to a Cleaning
Program. If (and ONLY if) a leak reoccurs
DO NOT
turn the machine off but instead press ‘Escape’ then
‘CONFIRM’ to Interrupt the program!
f)
(page 156).
CLEAN- 1.0.1 ISOLATE RESTRICTION
a) If the problem was occurring in Heat Disinfect return to it. If not place the machine in RINSE.
b) TWO (2) checks:
CHECK #1:
Figure right, water from the Fluid Sample port = UF Check
Valve #64 is bad!
CHECK #2:
No ‘squirts’ into the bucket = UF Check Valve #63 is bad!
c) Allow up to four (4) minutes
OR
until if a leak reoccurs. TWO (2) possible
scenarios below:
1)
IF (and ONLY if) NO leaks!
Proceed to
page 157
, procedure number
2)
IF a leak reoccurs!
With the “Interrupted” banner up, see procedure number
CLEAN- 1.0.3 LEAK OCCURS
a) Call debug screen 0 (Figure
right). If debug does not
appear press ‘Escape’ then
call screen 0.
b) Ignoring
Flow Error
watch
the
2
nd
window down
,
Valve Error
for one (1)
minute. TWO (2) possible
scenarios:
1)
IF (and ONLY if) Valve Error = 0 OR ‘blinks to 1’ for LESS THAN two (2) seconds:
proceed
to
page 562
, procedure number
2)
IF (and ONLY if) Valve Error EVER = 1 MORE THAN (2) seconds:
Proceed to
page 189
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...