2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 302
T- 7.0.0 ISOLATE TEMPERATURE SENSOR
These are non-routine TROUBLESHOOTING procedures! Follow them exactly to avoid error!
WARNING! Hydraulic alarms are not announced in Service Mode!
A) Figure right, ENSURE distribution board position #4,
“PH-PR”, remains VACANT!
B) ENSURE Cond Cell #7 is plugged into distribution board
position “X7, COND”!
C) Figure right, the
FOUR-
RESISTOR SET
is required.
D)
Enter Service Mode → Calibrate
Sensors → Temp Sensor.
E) Proceed to
STEP #1
(page 302).
STEP #1
a) The screen should say
“Connect a 6.808 K ohm resistor…”
b) Figure right, avoiding VACANT position #4, place the 34°
C
(6.808 KΩ) plug into the
2
nd
distribution board position
from the left, “MON-NTC”.
c) Is the screen’s [
Pre-Temperature Reference]
between
64 and 76 (Yes or No)?
Yes
Between 64 and 76! ‘Sharply’ press ‘CONFIRM’. If no Error banners
1
appear see
STEP #2
1
If an “Operator Error” OR
”Actuator Board Error”
banner appears proceed
to
page 303
, procedure number
No NOT between 64 and 76! ENSURE the 34° C plug is placed properly at the
2
nd
position from the
left! If okay, see procedure number
(page 303).
STEP #2
a) The screen should say
“Connect a 5.117 K ohm resistor...”
.
b) Place the 41° C
(5.117 KΩ) plug into the
2
nd
distribution board position from the left. Is
[Pre-
Temperature Reference]
between 157 and 169?
Yes
Between 157 and 169! ‘Sharply’ press ‘CONFIRM’ then see
STEP #3
(page 303).
No NOT between 157 and 169! ENSURE the 41° C plug is placed properly at the
2
nd
position from
the left! If okay, see procedure number
(page 303).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...