2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 81
F- 7.4.6 ISOLATE FLOW PUMP INPUT
a) ENSURE “Select Program” remains up!
b)
Return the dialyzer connector to the shunt
door!
c)
Return valve #35 to valve #30’s distribution
board position!
d) Figure right, remove the Flow Pump’s clear INPUT
tubing and measure from it for one (1) minute.
800
ml or more collected?
Yes
800 ml OR more! See procedure number
(page 81).
No
Less than 800 ml! See parts a THROUGH c below:
a) Reconnect the drain AND Flow Pump tubing.
b) Return valves #30 AND #35 to their distribution board positions.
c) TWO (2) checks below:
CHECK #1
Per the Figure below, check for tubing restrictions to and from post-Cell #13.
CHECK #2
There may be a restriction inside Air Removal Chamber #69!
F- 7.4.6.1 800 ML OR MORE
a) Reconnect the drain and Flow Pump tubing.
b) Return valves #30 AND #35 to their distribution board positions.
c) Plug BOTH concentrate connectors into their rinse ports!
(page 82).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...