2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 128
1)
IF (and ONLY if) PDIA AND ADIA REMAIN between 2.0 and 8.0:
2)
IF PDIA OR ADIA does DOES NOT REMAIN between 2.0 and 8.0:
See parts A THROUGH C
below:
A) A procedure in a different Section is performed next.
NOTE
this page and procedure number
(F- 18.2.0) because you may be prompted to return here.
B) BEFORE continuing to part C, proceed to
page 527
, to perform
C) If a leaking balancing chamber valve was NOT located in part B, see procedure number
F- 19.0.0 TROUBLESHOOT BAD LOADING PRESSURE
a) Return to Dialysis Program (
“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
b) The Deaeration motor shaft MUST be rotating COUNTERCLOCKWISE!
c) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM!’
d) WITHOUT LOOKING AWAY, simultaneously watch for a “No Water” alarm and Loading Pressure for five
(5) minutes. Is Loading Pressure now cycling between a PEAK of somewhere between 23 and 27 psi
and low of NEVER LESS THAN 11 psi and (Yes or No)?
Yes
Between 23 and 27 psi! Assuming a Flow Error does not occur perform an acid clean! If
(and ONLY if) the Loading Pressure problem reoccurs in the near future see procedure
number
(page 128).
No Cycles to less than 23 psi! See procedure number
(page 128).
F- 19.1.0 PRESSURE TEST HYDROCHAMBER
a) A procedure, in a different Section, is performed next.
NOTE
this page and procedure number (F-
19.1.0) because you may prompted to return to here.
b) BEFORE continuing to part c, proceed to
page 139
c) If the Hydrochamber leak was not located in part b return to Dialysis Program
(“Select Program” →
‘Dialysis’ → ‘CONFIRM)!
d) The Deaeration motor shaft MUST be rotating COUNTERCLOCKWISE!
e) WITHOUT LOOKING AWAY simultaneously watch for a “No Water” alarm AND Loading Pressure, for
five (5) minutes. Is Loading Pressure now cycling between a PEAK of somewhere between 23 and 27
psi and low of NEVER LESS THAN 11 psi and (Yes or No)?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...