2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 427
FIL- 3.5.0 -250 MMHG HELD / ISOLATE VALVE #43 FLOW
a)
IMPORTANT! Return the dialyzer lines to the
shunt and close the door!
b)
IMPORTANT! Turn Dialysate Flow on (Flow on/off
lamp STOPS blinking)!
c) Figure right, separate the Air Sensor Connector’s brown and blue
wires.
d) Obtain a 1000 ml (or larger) graduated cylinder!
e) Call debug screen 1.
Allow FILACT = 1!
f) Figure right, if a ‘Quick Connector’ is used, at the end
of the ‘to drain’ tubing an
adaptor
is required!
g) Measure drain flow for two (2) minutes. TWO (2)
possible scenarios:
1)
IF (and ONLY if) 600 ml or more collected:
2)
IF less than 600 ml collected:
ENSURING Dialysate Flow was on, perform parts a AND b below:
a)
Figure below, CAREFULLY trace the wire harness from distribution board positions “X9, P-
Dial” AND “X10, CFS” to ENSURE they terminate at the correct Pressure Transducer.
If NOT,
this may be the problem!
b)
Reconnect the drain then proceed to
page 41
procedure number