2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 47
F- 5.0.0 FLOW ERROR ‘1’ OR QDL NOT BETWEEN 720 AND 880 OR BC SWITCH > 400
a)
IMPORTANT!
CLOSE THE SHUNT DOOR!
b) The following is ESPECIALLY IMPORTANT if the machine was
worked on by someone who may have connected the BALANCING
CHAMBER VALVES (#31 through #38) incorrectly between the
distribution board and their solenoid terminals (Figure right)!
c) TWO (2) possible scenarios 1) or 2) below:
1)
IF (and ONLY if) SURE the machine was NOT worked on previously for this Flow problem:
(page 48).
2)
IF the machine WAS or MAY have been worked on:
See parts a THROUGH c below:
a) Leave the machine in Dialysis Program and
DO NOT
turn Dialysate Flow “OFF”!
b)
, without unplugging valves from the distribution board,
CAREFULLY
trace the wires from EACH valve to its distribution board position.
c) If (and ONLY if) a wiring error is NOT located see procedure number
(page 48). If a
wiring error was located this may be the problem.
Figure 9 – Distribution Board / Balancing Chamber Valves