2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 71
F- 7.0.5 ISOLATE VALVE #26 / VALVE #25 ACTUATOR-TEST BOARD CONTROL
a)
ENSURE “Select Program” REMAINS up!
b) EXCEPT for valve #37 and valve #30, ENSURE ALL OTHER valves are plugged into their PROPER
distribution board positions!
c) TWO (2) ‘touch’ checks on valve #26 AND #25’s (black) solenoids:
TOUCH CHECK #1:
solenoid.
TOUCH CHECK #2:
next page
), above (blue) Dialysate Pressure Transducer #9, is
Valve #25
. Touch its solenoid then see part d.
below, respond according to BOTH ‘touch’ checks:
Table J – Valve #26 / #25
Valve #26’s
Solenoid
Valve #25’s
Solenoid
Your Response
Warm (or Hot)
Warm (or Hot)
(page 72)
Cold
Does not matter
NOTE
ONLY
VALVE #26 will be checked and proceed
to
page 192
Does not matter
Cold
NOTE
ONLY
VALVE #25 will be checked and proceed
to
page 192
Figure 15 – Valve #26