2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 519
E) Per the Figure below AND as directed in
below, measure from the solder (back) side of the
Sensor Board’s X2 connector starting at pin #2:
Figure 90 – Electronic Card Cage (Sensor Board)
Table 10 – Sensor Board ‘X2’ DC Voltage Checks (Refer also to
X2 Pin
X2 Pin Location
Your Response:
2
Top row, first pin
from the REAR of
the machine
•
IF less than 0.5 volts DC:
Measure at pin 6 (next row in table)
•
IF more than 0.5 volts DC:
Bad Actuator-Test Board (see page 9).
6
Top row, third pin
from the REAR of the
machine
•
IF less than 4.5 volts DC:
•
IF more than 5.2 volts DC:
(page 520).
•
IF between 4.5 and 5.2 volts DC:
From debug screen 4,
does
LEAK
= the
pin 6 measurement /- 0.3 volts DC?
Yes
LEAK
= pin 6 (+/- 0.3). Measure at pin 7 (next row in table).
No
LEAK
NOT /- 0.3 of pin 6 = Bad Sensor Board (see page 9).
7
Bottom row, fourth
pin from the REAR of
the machine
•
IF less than 4.0 volts DC:
•
IF more than 6.0 volts DC:
(page 520).
•
IF between 4.0 and 6.0 volts DC:
From
debug screen 4, does
DIMN
= the
pin 7 measurement /- 0.3 volts DC?
Yes
DIMN
= pin 7 (+/- 0.3). See procedure number
(page 520).
No
DIMN
does NOT /- 0.3 of pin 7= Bad Sensor Board (see page 9).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...