2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 309
T- 7.5.0 VERIFY MEASURED TEMPERATURE
At this point it does not matter if the external meter is 37.0° C! WITHOUT LOOKING AWAY, watch it for one
(1) minute noting its highest and lowest values. TWO (2) possible scenarios:
1)
IF (and ONLY if) temperature changes more than +/- 0.2° C:
ENSURING the external flow
indicator’s ‘bob’ is rising and falling see procedure number
(page 309).
2)
IF temperature does NOT change more than +/- 0.2° C: DO NOT
press ‘CONFIRM’ until instructed.
Based on the external meter’s reading, TWO (2) possible scenarios i) or ii) below:
i) IF (and ONLY if) between 36.9 and 37.1° C:
Proceed to
page 311
, procedure number
ii) IF NOT between 36.9 and 37.1° C:
See parts a through c below:
a) Select the
[TEMP DAC]
data box, it turns bright yellow.
b) Adjust
[TEMP DAC]
(+/- 2 = +/- 0.1
°
C) then sharply press ‘CONFIRM’.
[TEMP DAC]
MUST be
pale yellow/white (NOT GRAY)
to continue!
c) Allow
FIVE (5) FULL
minutes while watching the external flow indicator’s ‘bob’.
Does it EVER remain down for longer than four (4) seconds?.
Yes
‘Bob’ stays down longer than four seconds! See (ABOVE) procedure
number
page 306
No ‘Bob’ never stays down longer than four seconds! Repeat procedure number
(page 309).
T- 7.5.2 TEMPERATURE IS UNSTABLE
a) Continue to watch the meter for up to six (6) FULL minutes ENSURING the external flow indicator’s ‘bob’
is rising and falling. If the external meter does NOT fall to less than 33° C continue to part b. If it DOES
see (ABOVE) procedure number
page 308
).
(page 298) performed in THIS troubleshooting session?
Yes
T- 6.0.0 was already performed! See procedure number
(page 310).
No a) Turn the machine OFF!
b) Turn the machine on and return to Dialysis Program (“Select Program
” → ‘Dialysis’ →
‘CONFIRM’).
b) See (ABOVE) procedure number
page 298
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...