2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 263
T- 2.0.0 HEATER VOLTAGE LESS THAN 10.0 VOLTS AC
a) Remaining in Dialysis Program or Heat Disinfect, to prevent pulling cables loose,
GENTLY open the card
cage.
b)
Set your volt meter to DC voltage (V
DC
)!
c)
Connect the black lead to chassis ground (see
d)
CAUTION!
Signals will be measured at pins that are VERY close to others and touching them
together, with a standard meter lead,
WILL CAUSE DAMAGE
! As directed below, make your RED
meter lead PROTECTED!
DO NOT CONTINUE UNTIL YOU HAVE DONE THIS!
e) Figure below, at the top of the Power Logic board, closest to the screen, locate its twenty-pin X2 cable.
.
Figure 43 – Power Logic Cable X2 / Pin 6
f)
Measure from
the solder (rear) side of the X2 cable, at
pin 6
(TOP row, three pins from the REAR of the
machine). A good signal is 9.0 volts DC OR MORE?