2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 295
T- 5.0.0 CHECK TEMP (SCREEN 4)
From debug screen 4, is
TEMP
LESS THAN 1.0?
Yes
TEMP
is less than 1.0! See procedure number
No
TEMP
is more than 1.0! Proceed to
page 297
, procedure number
T- 5.0.1 TEMP LESS THAN 1.0
a)
Turn the machine OFF!
b) Open the card cage.
c) Figure right, now place the
274
Ω
from the
TWO-RESISTOR SET
in
the
2nd
distribution board position
from the left, “MON-NTC”.
d)
Set your CALIBRATED
volt meter to resistance (Ω)!
e)
Connect the meter’s black lead to chassis ground (see
below, measure at the solder side of Sensor Board’s ‘X2’ connector at
pin 12
(top row, 6
pins from the rear of the machine).
Figure 53 – NTC #3 Checks
g) Figure right, reading the meter’s numeric AND
units display LESS than 300
Ω
(Yes or No)?