2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 547
OVER- 5.2.0 HEAT DAMAGE CHECKS
a)
IMPORTANT!
Turn the machine OFF to prevent
electrocution!
b) Per the Figure right, remove the heater and check its
opening into the top of the hydrochamber for damage i.e.
melting. If (and ONLY if) damage is LOCATED replace
the hydrochamber. If (and ONLY if) damage is NOT
located continue to part c.
c) All systems that may cause overflow are checking okay.
Assuming all above procedures above were performed
correctly the Troubleshooting Guide cannot locate the
problem!
LEFT BLANK INTENTIONALLY