
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 204
Yes
The Balancing Chamber Valves cycle! Valve #36 OR its blue wire harness is bad.
No Valves DO NOT cycle! See procedure number
(page 204).
VE- 6.0.3 VALVE #36 UNPLUGGED AND VALVES DO NOT CYCLE
a) At the distribution board (Figure above), unplug
Valve #37’s
connector.
b) Watch the Balancing Chamber Valve ‘dots’ for forty-five (45) seconds. Do they begin to cycle between
white and blue?
Yes
The Balancing Chamber Valves cycle! Valve #37 OR its blue wire harness is bad.
No
The Balancing Chamber Valves DO NOT cycle after Valves #31, #34, #36, and #37
have all
been unplugged! See parts A and B below:
A
)
CAREFULLLY return ALL Valve connectors to the distribution board;
B) Turning the machine off in between, swap in each component (see the Component List
below), one at a time, returning to Dialysis Program in between, until
VERR
remains = 0
for ten (10) minutes indicating the last component swapped in is the problem.
Component List
1)
Actuator-Test Board;
2)
Actuator Cable; 3
)
Distribution board
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...