2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 59
F- 6.3.0 RESTRICTED DRAIN FLOW / ISOLATE ‘DRAIN’ VALVE #30 / ACTUATOR-TEST BOARD
Per the Figure below, touch valve #30’s (black) solenoid. Is it HOT?
Yes
Solenoid is hot! See procedure number
(page 59).
No
Solenoid is NOT hot!
NOTE
ONLY
VALVE #30 will be checked and proceed to
page 192
Figure 12 – Valve #30
F- 6.3.1 ISOLATE DRAIN VALVE #30
a) Turn Dialysate Flow OFF (Flow on/off lamp blinks).
b) Check through Valve #30’s input
AND
output tubing for bio-growth
restrictions.
c) Figure right, at the rear of the machine, remove the tubing from the
‘To Drain’ (
BOTTOM!
) nozzle.
d) Check inside the nozzle for restrictions!
e) Per the Figure above, remove the tubing from Valve #30’s INPUT nozzle.
f) Attach a syringe, with tubing attached that will fit SNUG over Valve #30’s nozzle.
g) Push or pull on the syringe plunger. If Valve #30 is open the plunger moves with no resistance. TWO (2)
possible scenarios next page: