2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 392
b) Return Post Dialyzer Cond Cell #13 connector to distribution board position labeled “x13,
COND-POS”.
c) From debug screen 0, ENSURE Post Dialyzer Temp reads more than 35° C AND Post Dialyzer
Cond MUST more than 10.000.
CO- 10.0.66 ISOLATE PRE SENSORS STABILITY
a) Figure right, remove Pre the Dialyzer Temp Sensor (NTC
#3’s) connector from the
2
nd
distribution board position from
the left, “MON-NTC”.
b) Using a flashlight, CAREFULLY check the vacant “x3”
distribution board position for corrosion and/or damaged
pins.
If damage is located this may be the problem!
c) Avoiding vacant position #4, place the
6.04 KΩ
plug, from the
TWO-RESISTOR SET
, into the
Pre Dialyzer Temp Sensor’s NTC #3 distribution board position “x3, MON-NTC.
d) Remove Pre Dialyzer Cond Cell #7’s connector from distribution board position “x7, COND”.
e) Using a flashlight, CAREFULLY check the vacant “x7” position for corrosion and/or damaged pins.
If
damage is located this may be the problem!
f)
Leaving the 6.04 KΩ plug installed, place the
274 Ω
resistor plug, from the
TWO-RESISTOR
SET
, into the Pre Dialyzer Cond Cell #7’s distribution board position, “x7, COND”.
g) From debug screen 5, TWO (2) checks then proceed according to the TWO (2) possible scenarios
after check #2.
CHECK #1 (TPRE
):
WITHOUT LOOKING AWAY, watch
TPRE
(3
rd
column from
left, at top) for exactly one minute noting its highest and lowest values. Subtract the
lowest value seen from the highest. Two (2) or less = GOOD! MORE THAN 2 = BAD.
CHECK #2 (CPRE):
Watch
CPRE
(4
th
column from right, second box down) for
exactly one minute noting its highest and lowest values Subtract the lowest value
seen from the highest. Eight (8) or less = GOOD! MORE THAN 8 = BAD.
1)
IF (AND ONLY IF) TPRE AND / OR CPRE are BAD:
Proceed to
page 394
procedure number
2)
IF (and ONLY if) TPRE AND CPRE are GOOD:
See parts a through c below.
a) Return Pre Dialyzer Temp Sensor NTC #3s connector to distribution board position “x3, MON-
NTC”. From debug screen 0, ENSURE Pre Dialyzer Temp reads more than 35° C.
b) Return Pre Dialyzer Cond Cell #7 connector to distribution board position “x7, COND”. From
debug screen 0, ENSURE Pre Dialyzer Cond reads more than 10.0.
(page 393).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...