2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 586
E)
Reset ALL alarms!
F) Select the screen’s lower
[Test & Options]
tab.
G) Select the ‘Diasafe Test’ button.
H) Press ‘CONFIRM’ to start the test then allow it to finish! TWO (2) possible scenarios:
1)
IF (and ONLY if) “Diasafe Test Passed”:
Repeat the
test once more as the failure may be intermittent! If it
passes again DO NOT continue!
2) IF “Diasafe Test Failed”:
a)
Call the Home screen.
b) Set
[UF GOAL]
to 1000 ml; Set
[UF Time]
to 1:00 hr.
c) Press ‘CONFIRM’.
d) Allow two (2) minutes BEFORE continuing.
e) Using a flashlight, ENSURE no air flowing through the external flow
indicator sight tube.
LEFT BLANK INTENTIONALLY