
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 335
CO- 1.0.9 VERIFY DIALYSATE LIMITS
a) Select the ‘Dialysate’ tab (Figure below).
Dialysate SCREEN
b)
TCD
is based on Dialysate Composition and indicates what the ‘Actual’ Conductivity would be if
everything was perfect.
c) Using [Alarm Width], spread Conductivity Limits as wide as they will go and press ‘CONFIRM’.
d) Using [Alarm Position], set the upper Limit to 0.5 above
TCD
and press ‘CONFIRM’.
e) Based on the ‘Actual’ [
Conductivity
] window (Figure right),
TWO (2) possible scenarios:
1)
IF (and ONLY if) [Conductivity] is STABLE between 13.4
and 14.3 mS:
(page 336).
2) ALL OTHERS:
Proceed to
page 337
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...