2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 170
CLEAN- 2.1.2 ISOLATE FLOAT SWITCH #5
a) This procedure requires up to
four (4) minutes
to prevent labor-intensive error.
b) Direct the tubing into the bucket for up to four (4) minutes. Flow will either:
1)
Continue at more than
130 ml per minute OR
2)
Stop. TWO (2) possible scenarios:
1)
IF (and ONLY if) flow continues at 130 ml or more every minute:
Return to (ABOVE)
(page 167).
2) IF flow STOPS:
TWO (2) possibilities:
1)
Bad Float Switch #5
*
OR;
2)
Bad Hydrochamber.
*
To LOCATE the Float Switch refer to
(page 129)
LEFT BLANK INTENTIONALLY