2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 129
Yes
Peak between 23 and 27 psi! Call debug screen 0 to watch
Flow Error
for three (3)
minutes. If it remains = 0 perform an ACID CLEAN. If (and ONLY if) the Loading Pressure
problem reoccurs in the near future FIVE (intermittent) possible bad components (see
Component List below). Swap in each, one at a time, and in between check if loading
pressure of between 23 and 27 psi can be achieved.
No
Peak IS NOT between 23 and 27 psi! FIVE (5) possible bad components (see Component
List below). Swap in each, one at a time, and in between, check if Loading Pressure PEAKS
to between 23 and 27 psi
can be achieved indicating the last component swapped in was the
problem.
COMPONENT LIST:
1)
Deaeration Pump head
*
;
2)
Deaeration Pump motor
*
;
3
) Loading Pressure Valve #65
**
;
4)
Float switch (see Figure below);
5)
Hydrochamber (see Figure below).
* To LOCATE these components refer to
(page 19)
**
Attempt to calibrate the new Valve #65 to achieve a peak loading pressure of between 23
and 25 psi
Figure 28 – Hydrochamber / Float Switch #5 / Heater
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...