2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 99
F- 9.0.42 MORE THAN 800 ML INPUT FLOW TO VALVE #26 / ISOLATE VALVE #24
a)
Return to Calibrate Hydraulics → Flow Pressure.
b) Press ‘CONFIRM’ then ENSURE the Flow Motor is rotating!
c) Remove the BLUE dialyzer connector from the shunt door.
d) Measure flow from the dialyzer connector for one (1) minute!
e) Press ‘CONFIRM’ to return to “Calibrate Hydraulics”.
f) More than 800 ml collected?
Yes
More than 800 ml! Good flow from Valve #24! See procedure number
No Less than 800 ml! See parts a THROUGH f below.
a)
Per the Figure below, remove Valve #24’s INPUT tubing!
b)
Direct it into the bucket.
c)
Return to Calibrate Hydraulics → Flow Pressure.
d)
Press ‘CONFIRM’ THEN ENSURE the Flow Motor is rotating.
e)
Measure from the tubing for one (1) minute.
f)
Press ‘CONFIRM’ to return to “Calibrate Hydraulics”.
g)
More than 800 ml collected?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...