2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 570
SECTION 22 - “WD: FAIL LONG PULSE” OR “WD: FAIL SHORT PULSE”
Turn the machine OFF for two (2) seconds, then back on. TWO (2) possible scenarios 1) or 2) below:
1)
IF “WD: Fail Long Pulse” OR “WD: Fail Short Pulse” does NOT reoccur:
Turn the machine off
and on ten (10) times allowing twenty (20) seconds in between as the problem may be intermittent! If
“WD” alarm reoccurs see scenario #2.
2)
IF “WD: Fail Long Pulse” OR “WD: Fail Short Pulse” REOCCURS:
Perform parts a THROUGH d
below:
a) Figure below, being
CAREFUL
to
NOT
pull the 24V Power Harness loose, open the card cage!
b)
IMPORTANT! Set your CALIBRATED volt meter DC voltage (V
DC
)
c)
Connect the meter’s ground (black) lead to chassis ground (see
page 4)).
d)
ENSURING
the
“WD: Fail Long Pulse” OR “WD: Fail Short Pulse”
banner
is on the screen,
(page 571).
LEFT BLANK INTENTIONALLY