2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 376
CO- 9.8.3 BALANCING CHAMBER VOLUME IS NOT FACTORY RECORDED
A clear, preferably glass, 100 ml graduated cylinder OR even MUCH
more preferable a calibrated scale is required!
DO NOT
use a larger
cylinder or gross error will occur!!
a)
WITHOUT turning flow off
, figure right, remove the tubing
from the ‘To Drain’ (*
BOTTOM!
) nozzle at the REAR OF THE
MACHINE. *
CAUTION!
Measuring from the end of the ‘to
drain’ tubing causes gross error!
b)
Allow ten (10) drain ‘pulses’ to ENSURE flow stability!
c)
ACCURATELY,
collect two (2) consecutive drain ‘pulses’
into the cylinder. If a scale is used 1 ml = 1 gram. If a
scale is not used read the bottom of the cylinder’s
meniscus curve (Figure right).
d) Reattach the drain tubing! TWO (2) possible scenarios:
1) IF (and ONLY if) measured volume is NOT
consistent between 58 and 62 ml:
Proceed to
page 535
, to perform
FOR A LEAKING BALANCING CHAMBER
DIAPHRAGM
2) IF measured volume is consistent between 58 and 62 ml:
See parts a AND b below:
a) Call debug screen 3.
BCV
(upper right,
B
alancing
C
hamber
V
olume) was
entered during the last BCV calibration (example:
603 = 60.3 ml)
b) Does
BCV
AND the measured volume match?
Yes
BCV
and the RECORDED volume match! See procedure number
No
BCV
and the RECORDED volume do NOT match! Read parts a THROUGH c below
before performing them:
a) Perform the BALANCING CHAMBER VOLUME CALIBRATION,
BUT instead
of measuring again enter your RECORDED volume.
b) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
c) Call debug screen 3. If the calibration took
BCV
= your RECORDED volume.
(page 377).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...