2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 414
FIL- 1.0.6 PRE AND POST COND NOT WITHIN 0.4 OF EACH OTHER
Figure below, to the right of Pre Dialyzer Cond Cell #7, is Post Dialyzer Cond Cell #13 present?
Yes
Cond Cell #13 present! See procedure number
(page 414).
No
Cond Cell #13 NOT present! Proceed to
page 416
, procedure number
Figure 66 – OLC
Pre AND Post Components
FIL- 1.0.7 ISOLATE OLC SENSORS CONNECTIONS AND CABLES
a) Per the Figure above,
FOUR (4) IMPORTANT CHECKS
:
CHECK #1:
ENSURE the Sensor ribbon cable is plugged in PROPERLY!
CHECK #2:
ENSURE the cable from position “x13, COND-POS” terminates at Post
Conductivity Cell
#13!
CHECK #3:
ENSURE the cable from position “x44, NTC-POS” terminates at Post Temp Sensor #44!
CHECK #4:
ENSURE ALL cables show no signs of insulation damage!
b) Were the cables OKAY?