2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 306
T- 7.2.2 [TEMP DAC] BETWEEN 120 AND 140 / VERIFY FLOW
The external flow indicator’s ‘bob’
is watched to see if it EVER, even once, stays down longer than four (4)
seconds AND the external meter is watched to see if it EVER, even once, exceeds 38.9° C. WITHOUT
LOOKING AWAY watch ‘bob’ AND
the meter for five (5) minutes. TWO (2) possible scenarios:
1)
IF (and ONLY if) ‘bob’ NEVER stays down longer than four (4) seconds:
Proceed to
page 308
2) IF ‘bob’, even once, stays down longer than four (4) seconds:
T- 7.2.4 ‘BOB’ DOWN LONGER THAN FOUR SECONDS / ISOLATE INTERNAL ALARM LIMIT
a) Figure right, remove the connector from the
2
nd
distribution
board position from the left, “MON-NTC”. This is temperature
sensor NTC #3.
b) The screen’s
[Monitor Reference]
window should be less
than 30°C if NTC #3 was unplugged!
c) Does ‘bob’ start to move up and down?
Yes
‘Bob’ moving! Leaving NTC #3 unplugged, see procedure number
(page 306).
No ‘Bob’ not moving!
A)
Return NTC #3’s connector to
2
nd
position from the left, “MON-NTC”;
B)
There is an unannounced “No Water” OR Flow Error present;
C)
Turn the machine off;
D)
Turn the machine back on and return to Dialysis Program
(“Select Program” → ‘Dialysis’ →
‘CONFIRM’) to troubleshoot these problems!
T- 7.2.5 ‘BOB’ WAS NOT RISING BUT IS NOW / ISOLATE POSSIBLE HIGH TEMPERATURE
TO PREVENT DAMAGE,
WITHOUT LOOKING AWAY, watch the external meter for six (6) minutes or until
if it, even if only once, exceeds 38.9° C! TWO (2) possible scenarios:
1)
IF (and ONLY if) the meter NEVER reaches 38.9° C:
(page 307).
2)
IF the meter DOES reach 38.9° C at least ONCE:
See parts a THROUGH d below:
a) Turn the machine OFF!
b)
RETURN NTC #3’s connector to the
2
nd
distribution board position from the left, “MON-
NTC”
.
c) Using a temperature meter measure incoming (RO) water temperature. If (and ONLY if) less than
37.0º C see part d. If more than 37.0° C incoming water is too high and this may be the problem!
d) Return to Dialysis Program (
“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
e) Proceed to
page 241
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...