2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 423
FIL- 2.4.0 FILACT = 0 / ISOLATE POSSIBLE FLOW PROBLEM
a) Call debug screen 6 to locate
BC Switch
(middle column). If EVER = 897 or more,
even just once, indicates a masked Flow Error! In any event see part b.
b) WITHOUT LOOKING AWAY, watch
BC Switch
for two (2) minutes. TWO (2) possible scenarios:
1)
IF (and ONLY if) BC Switch NEVER, EVER = 897 or more!
2) IF BC Switch = 897 or more, even just once!
Proceed to
page 20
FIL- 2.4.2 BC SWITCH NEVER = 897 / ISOLATE X6
a)
Figure right
, return the Air Sensor’s
connector to the distribution board. If CBE
equipped, using a flashlight, ENSURE the
connector covers ALL CBE board pins,
especially the top pin!
b) Allow thirty (30) seconds as
FILACT
response is
not instantaneous!
c) Call debug screen 1. WITHOUT LOOKING AWAY, watch
FILACT
for two (2) minutes or until if it EVER
= 1. TWO (2) possible scenarios:
1)
IF (and ONLY if) FILACT = 0 ALWAYS:
Proceed to
page 425
, procedure number
2) IF FILACT EVER = 1:
See parts a THROUGH e below:
a)
ENSURE
the female Air Sensor’s Connector is plugged in PROPERLY! If CBE equipped using
a flashlight,
ENSURE
the connector covers ALL CBE board pins, especially the top pin!
b)
WITHOUT pulling on the wires
, unplug the blue and
brown wires from the probes at the top of Chamber
#69.
c) Figure right, connect the wires together.
d) Allow thirty (30) seconds.
e) WITHOUT LOOKING AWAY, watch
FILACT
for one (1)
minute! ALWAYS = 0?
Yes
FILACT ALWAYS
= 0! Leaving the wires connected together proceed to
page 425
No
FILACT
= 1 at least once! See procedure number
(page 424)
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...