2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 224
T- 1.0.122 TEMP SENSOR CALIBRATION SUCESSFUL / VERIFY TEMPERATURE ALARM
a) Return NTC #3’s connector to the
2
nd
distribution board position from the left, “MON-NTC”.
b) Turn the machine OFF.
c) Turn the machine on and return to Dialysis Program (“Select Program”
→ ‘Dialysis’ → ‘CONFIRM’)!
d) From the Home screen, select the [Dialysate Flow] window.
e) Set the [Dialysate Flow] to 500 ml/min and press ‘CONFIRM’.
f) Allow five (5) minutes BEFORE continuing to part g.
g) ENSURING [
Temperature
] is between 35.1 and 38.9° C, AND is not
changing more than 0.2° C per minute, what color is the
[
Temperature
] window now? TWO (2) possible scenarios:
1)
IF (and ONLY if) pale yellow / white:
(page 225).
2) IF RED: Read this procedure before performing it.
Swap in the listed components (see
COMPONENT LIST
below), one at a time, with known good, and in between return to (ABOVE)
(page 222) to test each new component. If you return here swap in
the next component until the [
Temperature
]
window is pale yellow/white!
COMPONENT LIST: 1)
Actuator-Test Board
a
;
2)
Functional Board
a,b
.
a
To LOCATE these boards refer to
(page 9)
b
With the Functional board, to prevent “Cond Offset Failure”, place the
machine into
T and C Mode
(refer to
OPERATING MODES
, page
Error!
Bookmark not defined.
))
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...