2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 104
F- 11.0.0 ACFS = BETWEEN 0 AND 3.0
a) Turn Dialysate Flow OFF (Flow on/off lamp blinks)!
b) Press and release the ‘Reset’ key then immediately press and hold it for three (3) seconds. Allow thirty
(30) seconds. If a TMP occurs attempt RESET up to FIVE (5) times BEFORE continuing to part c.
c) Call debug screen 10. Is
ACFS
now = 3.5 or more?
Yes
ACFS
= 3.5 or more! See procedure number
No
ACFS
IS NOT 3.5 or more! See parts a THROUGH c below:
a) Figure right, unplug the CFS (#10)
transducer’s connector from
distribution board position “x10, CFS”.
b)
IMPORTANT!
To avoid error,
using
the screen’s clock (upper right),
allow
up to five (5) minutes as
ACFS
response IS NOT instantaneous!
c) After no more than five (5) minutes,
ACFS
should increase to 9.0 or more?
Yes
ACFS
= 9.0 or more!
A)
Return the
CFS transducer #10 to distribution board
position “x10, CFS”;
B)
See procedure number
(page 104).
No
ACFS
DOES NOT = 9.0 or more! Leaving the CFS transducer unplugged for now,
FOUR (4) possible bad components:
1)
Actuator-Test Board
a
OR;
2)
Sensor
Board
b
OR;
3)
Sensor Board cable OR;
4)
Distribution board.
a
A)
With the power off, swap in a known good Actuator-Test Board* (*to
LOCATE the board refer to Figure 4A, page 9);
B)
Return to Dialysis Program
(“Select Program” → ‘Dialysis’ →
‘
CONFIRM’);
C)
Allow up to
five (5) minutes
;
D)
If
ACFS
goes to more than 9.0 the previous Actuator-Test Board is bad.
b
A)
With the power off, swap in a known good Sensor Board (to LOCATE the
board refer to Figure 4A, page 9);
B)
Place the machine into
T and C Mode
(refer to
OPERATING MODES
(page
Error! Bookmark not defined.
));
C)
Return to Dialysis Program;
D)
Allow up to
five (5) minutes
;
E)
If
ACFS
goes to more than 9.0 the previous
Sensor Board is bad.
F- 11.0.4 CHECK DIALYSATE PRESSURE
a) Leaving Dialysate Flow OFF, remove both DIALYZER connectors from
the shunt and hold them at the level of the shunt door.
b)
Close the shunt door!
c) Call debug screen 0. Is the
PDlal
window (Figure right) between
-50 and +50?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...