2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 215
T- 1.0.0 IN DIALYSIS PROGRAM / TROUBLESHOOTING TEMPERATURE
a) Select the [
Temperature
] window and set it to 37.0° C!
b)
Press ‘CONFIRM’.
c) Call debug screen 1. TWO (2) checks:
Check #1: 5V EST (right column):
Between 4.7 and 5.3?
Check #2: 12V EST (right column):
Between 11.7 and 12.3?
d) TWO (2) possible scenarios:
1)
IF (and ONLY if) BOTH are in range AND stable i.e. not changing more than 0.1:
See
2) IF one or both is NOT REMAINING in range OR is unstable:
See parts A THROUGH D below:
A) Turn the machine OFF!
B) Using ESD precautions, reseat all card cage circuit boards.
C) Turn the machine on and r
eturn to Dialysis Program (“Select Program” → ‘Dialysis’ →
‘CONFIRM’).
D) Call debug screen 1. If (and ONLY if)
5V EST
and/or
12 EST
are still not in range OR unstable
proceed to
page 597
, procedure number
5V EST
and/or
12 EST
are in range see
T- 1.0.1 ISOLATE FLOW ERROR
a) If the Automated Tests are running (screen reads “Test:.…”) allow them to finish.
b) Remove the ‘dummy venous chamber’ from the Level Detector.
c)
Do NOT reset alarms!
d) Call debug screen 0 to locate
Flow Error
. If it
EVER
= ‘1’,
even just once, indicates a masked Flow Error.
e) WITHOUT LOOKING AWAY, watch
Flow Error
for one (1) minute. Is it EVER = 1?
Yes
Flow Error
= 1, even if only once! Proceed to
page 20
No
Flow Error
ALWAYS = 0! See parts a THROUGH c below:
a) Return the concentrate connnectors to their rinse ports.
Parts b and c next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...