2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 270
T- 2.1.4 DIODE 17 CHECKED GOOD OR POWER CONTROL BOARD WAS REPLACED
a)
IMPORTANT!
Figure below, ENSURE all cables are PROPERLY connected to the Power Control
board!
NOTE:
A good diode 17 does NOT necessarily mean the Power Control board is good
b) If troubleshooting a DIAL
YSIS PROGRAM temperature problem return to it (“Select Program” →
‘Dialysis’ → ‘CONFIRM’). If a HEAT DISINFECT problem return to HEAT DISINFECT.
c)
Set your meter AC voltage (~ V, V
AC
)
.
d) Figure right, measure BETWEEN the Heater Connector’s
BROWN and BLUE wires. TWO (2) possible scenarios:
1)
IF more than 100.0 volts AC!
Return
NTC #2’s
connector to the
1st
distribution board position from
the left, “CON-NTC”. The new boards fixed the
problem!
2)
IF less than 100.0 volts AC!
See procedure number
(page 270).
T- 2.1.5 HEATER VOLTAGE LESS THAN 10 VOLTS AC
a)
Set your meter to DC voltage (V
DC
).
b)
Connect the black lead to chassis ground!
c) Measure again at the Power Logic board’s
T5
TOP pin. If necessary, refer to
(page 264).
d) LESS THAN two (2.0) volts DC?
Yes
T5 LESS THAN 2.0 volts! Proceed to
page 273
, procedure number
No T5 between 2.0 and 28.0 volts! See procedure number
(page 271).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...