2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 52
F- 6.1.1 PDIA NOT BETWEEEN 1.0 AND 8.5 OR UNSTABLE
a) Momentarily plug the RED (acid) concentrate connector into its rinse port to call “Select Program”.
b) Return the RED connector to acid concentrate!
c) Select the ‘Dialysis’ button but
DO NOT
press ‘CONFIRM’!
d) Remove the red DIALYZER connector from the shunt door and
HOLD IT as seen in the Figure right!
e) Continuous flow, more than 0.2 ml every minute, from the connector?
Yes
Continuous flow! Turn the machine off THEN proceed to
page 527
SECTION 18A – DIAGNOSTIC VALVE LEAK TESTS
No Flow! See parts a THROUGH c below:
a)
Return the connector to the shunt and close the door!
b)
ENSURE the TOP balancing chamber valves are REMAINING BLUE!
c)
From debug screen 0, Figure right, watch valve #41’s ‘dot’ for thirty
(30) seconds. TWO (2) possible scenarios:
1)
IF (and ONLY if) valve #41’s ‘dot’ cycles between white and blue:
Turn the
machine off THEN proceed to
page 527
SECTION 18A – DIAGNOSTIC VALVE
2) IF valve #41’s ‘dot’ STAYS white:
Call debug screen 4. Watch
PDIA
AND
ADIA
again for one (1) minute, TWO (2) possible scenarios i) or ii) below:
i)
IF (and ONLY if) PDIA AND / OR ADIA is NOT between 3.0 and 7.5 OR
changes more than 0.1 i.e. unstable:
Proceed to
page 462
, procedure
ii) IF PDIA AND ADIA are between 3.0 and 7.5 and DOES NOT change more
than 0.1:
See
(page 53).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...